Nanotechnology/Electron microscopy

Navigate
<< Prev: Optical Methods
>< Main: Nanotechnology
>> Next: Scanning Probe Microscopy

Electron microscopy

edit

An overview:

Electron microscopes uses electrons instead of photons, because electrons have a much shorter wavelength than photons and so allows you to observe matter with atomic resolution.

There are two general types of electron microscopes: the Scanning Electron Microscope (SEM) that scans an electron beam over the surface of an object and measures how many electrons are scattered back, and the Transmission Electron Microscope (TEM) that shoots electrons through the sample and measures how the electron beam changes because it is scattered in the sample.

 
A very simple sketch of a Transmission Electron Microscope (TEM) and Scanning Electron Microscope (SEM) compared to an optical transmission microscope and a cathode ray tube (CRT) TV screen - both systems have many things on common with the electron microscope. The optical microscope uses lenses to control the lights pathway through the system and is in many ways built up like a TEM - only the TEM uses electromagnetic lenses to direct the beam of electrons. The CRT uses electromagnetic lenses as the TEM and SEM to control the electron beam, and generates an image for the viewer by scanning the beam over a fluorescent screen - in the same way the a SEM generates an image by scanning the electron beam over a small sample.

Using electron beams however requires working in a vacuum environment, and this makes the instruments considerably larger and expensive. All electron microscope work under at least low pressures and usually in high vacuum chambers to avoid scattering the electrons in the gas. In environmental electron microscopes, differential pumping systems are used to actually have gasses present by the sample together with the electron beam.

Introduction to Electron Microscopy

edit

For imaging of nanoscale objects, optical microscopy has limited resolution since the objects are often much smaller than the wavelength of light. The achievable resolution for a wavelength   is often given by the diffraction limit   as

  (r.g., diffraction limit)

with numerical aperture  , which can be approximated by the largest angle of incidence   of the wavefront towards the sample,  .

Since   for the present purposes, we can approximate   and hence   where   is the radius of the objective lens aperture and   the working distance.

Optical microscopes can often reach a resolution of about   nm. For nanoscale resolution this is unfortunately not sufficient to distinguish for instance a single nanotube from two adhering to each other, since they have diameters of less than 100 nm.

The figure below gives an overview typical magnifications achievable by the different electron microscopes compared to a light microscope.

 
The different methods for microscopy cover a range of magnification roughly indicated by the bars in the figure. The resolution of optical microscopy is limited to about 200 nm. a) SEM image of the head of an ant facing a microfabricated chip with a pair of microfabricated grippers. The grippers are barely visibly at the tip of the arrow. b) SEM image of a gripper approaching a large bundle of carbon nanotubes. c) Closeup in SEM of the gripper and nanotubes. d) TEM image of a carbon nanotube suspended between two grippers. e) TEM closeup of the shells of carbon atoms in a carbon nanotube. On the nanometer scale this particular carbon nanotube does not show a well defined carbon shell structure.

Electron optical systems use electrical and magnetic fields to control the electron beam. Although the law of refraction in optics is exchanged with the Lorentz force in electrodynamics, the electron optical system has similar diffraction limits as optical systems, since they depend on the wave nature of the electron beam.

One can achieve a considerable improvement in resolution with instruments such as the transmission electron microscope and the scanning electron microscope that use electrons with De Broglie wavelength much smaller than that of visible light. The De Broglie wavelength λ of an electron with momentum p is

  (Eq. De Broglie wavelength)

where   is Plancks constant. The electron has rest mass   and energy  .

If an electron with charge   is accelerated from rest by an electrical potential  , to the electron beam energy  , it will have a wavelength of 1 nm at 1 eV decreasing to 1 pm at 100 keV where it will be travelling with 50% the speed of light.

This chapter will briefly review fundamental issues for electron microscopy that are similar for SEM and TEM: the limitations imposed by the electron optical beam system in the microscope column; the interaction of the electron beam with the sample; the standard image formation method in SEM and TEM. These issues are essential to understand the results and limitations reached in SEM and TEM microscopy.

For further details, please refer to reviews of electron microscopes and their applications, such as Goldstein et al. [1] that contains a thorough review of SEM, while Goodhew and Humphreys [2] is a more general introduction to both SEM and TEM.

The Electron Optical System

edit

For high resolution imaging, a well focused beam is required, just as in optical microscopy. Due to the short wavelength of electron beams with keV energies, as given by the #Eq de Broglie wavelength, the properties of the electron optical system and the electron emitter mainly defines the limits on the achievable beam diameter. The current density in the electron beam can be approximated by a Gaussian distribution of current density j [A/m²] as function of radius, r, from the beam center

 

with radius determined by  , giving a the full width half maximum  . Integrating   gives the total beam current

 

The electron optics impose a limit on the achievable beam current density and radius by the brightness of the electron emitter  , which is conserved throughout the system [3].

Brightness, ß, is a measure of the current per area normal to the beam direction and per element of solid angle [4]. At the center of the Gaussian beam,

 

and the brightness is related to the current density in eq SEM Gaussian beam profile. The emitter brightness   is determined by the type of electron emitter and the beam energy   [5]

 

with emission current density for W-filament sources about  ~3 A/cm², for LaB6 sources about 100 A/cm², while field emission guns (FEG) can reach 105A/cm². The energy spread of the electrons from the sources are about ΔE~1 eV and slightly lower for FEGs. Due to conservation of the brightness in the system, the beam diameter depends on current as

 

The ideal beam probe size determined by the conservation of brightness cannot be obtained in a real system. Effects such as aberration will make the minimum achievable beam diameter larger. Equation #eq SEM beam diameter however seem to adequately describe the beam diameter for the present discussion. Apart from the additional beam widening contributions, the image detection method imposes limits on useful values for the parameters in Eq. SEM beam diameter which differ for SEM and TEM.

Electron Range

edit

The electron optical system sets limitations to the achievable primary beam current and radius. The expected image resolution set by the primary beam cannot be reached if the signal detected for imaging is caused by electrons scattered far in the sample. The trajectory of an electron penetrating a bulk solid is a complex trajectory due to multiple elastic and inelastic collision events. As the primary electron (PE) penetrates into the sample it will gradually change direction and loose energy in collisions. The mean free path due to elastic and inelastic collisions,  , depends on the atomic number of the material and the PE energy. At 100 keV   for carbon and 5 nm for gold [6]. For samples thinner than   the main part of the PE will pass relatively unaffected through the sample, which is the basis for TEM.

 
Overview of electron electron scattering processes in bulk and tip-shaped specimens. The PE are scattered within the interaction volume, defined the electron range in the material. The range is longer than the mean free path  . The SE have a very short range, and only those created within that range from the surface can escape the material. This defines the SE escape depth.

SEM can be used for thicker specimens. The electrons that escape from the sample in a new direction compared to the PE due to elastic collisions are called backscattered electrons (BSE).

For samples thicker than  , the volume interacting with the scattered PE defines the range of the electrons in the material, and this is considerably larger than the minimum achievable primary beam diameters.

The electron range is about 1 µm at 10 keV for carbon, decreasing with higher atomic number for the material. Both the high energy PE and BSE generate secondary electrons (SE) by inelastic scattering events. The SE are generally defined as having energy below 50 eV while the BSE have energies up to the PE energy. The range of SE is typically 1 nm for metals and about 10 nm for insulators [7].

The short range of the SE make the yield of SE highly dependent on the energy lost by the PE within the SE range from the surface, and this makes high Z substances efficient generators of SE. The main emission of SE takes place in the region where the PE strikes the surface and within the SE escape depth from this region.

 
The electron range increases with beam energy. The internal structure of the EEBD deposits can be examined at high electron beam energies in SEM. At 5 kV with shallow penetration depth, the surface of the tips is clearly visible while at higher energies a core of more dense material becomes increasingly visible. At 100 keV and above, TEM images can achieve atomic resolution where the lattice planes in nanocrystals such as the gold nanocrystal in (c). The gold crystal is embedded in amorphous carbon with no clear lattice pattern.

Scanning electron microscopy (SEM)

edit

In a scanning electron microscope a beam is scanned over the sample surface in a raster pattern while a signal is recorded from electron detectors for SE or BSE. The PE energy is kept relatively low (1-30 keV) to limit the interaction volume in the specimen that will contribute to the detected signal. Especially low energy PE will provide high sensitivity to surface composition as they cannot penetrate far into the sample.

The figure above showed the effect of PE penetration depth of a carbonaceous nanostructure with a gold core, where only the surface is visible at low PE energies, while the carbon becomes increasingly transparent and the core visible at high PE energies.

The low energy SE can easily be attracted and collected by a positively charged detector and are hence an efficient source for an image signal. The standard SE detector is an Everhart-Thornley (ET) detector where a positively charged grid attracts the SE and accelerates them to sufficiently high energies to create a light pulse when striking a scintillator. The light pulse is then amplified by a photomultiplier. Despite the complex construction, the ET detector is remarkably efficient, but requires large   for effective collection of the SE by the charged grid.

Another SEM detector is the in-lens detector, where SE passing through the column aperture are accelerated towards a solid state detector. The in-lens detector complements the ET by being more efficient at short  .


Environmental SEM (ESEM)

edit
 
Simple sketch of an Environmental Scanning Electron Microscope (ESEM), where a differential pumping system with two pressure limiting apertures between the ultra high vacuum SEM column and the low vacuum sample chamber allows high pressures up to 10 hPa around the sample. This is enough to have liquid water at moderate cooling of 5 deg. C.

The ESEM makes it possible to use various gasses in the sample chamber of the microscope since there are narrow apertures between the sample chamber and the gun column, and a region in between that is connected to a differential pumping system. Pressures up to about 10 Torr are normally possible in the sample chamber.

The standard Everly-Thornhart SE detector would not work under such conditions since it would create a discharge in the low pressure gas. Instead a "gaseous secondary electron detector (GSD)" is used, as shown in the figure below. The GSD measures the current of a weak cascade discharge in the gas, which is seeded by the emission of electrons from the sample.

 
Two examples of images from an ESEM. Taken with a Philips XL-30 FEG. The first shows a electron beam deposited nanowire between two microelectrodes that has burnt after sustaining a high bias current. The other shows a multiwall carbon nanotube sample. Shorter working distances often improves image quality and so does a low beam current but it also increases the image acquisition time

In the ESEM one can work with for instance water vapour or argon as the environmental gas, and it is possible to have liquid samples in the chamber if the sample stage is cooled sufficiently to condense water.

Transmission electron microscopy (TEM)

edit


 
A Philips EM 430 TEM

When the specimen thickness is about the mean free path,  , TEM can be used to achieve high resolution images such as the image above where the atomic lattice of a gold nanocrystal is visible. Since the detected electrons are transmitted PE where the energy can be in the 100 keV range, the resolution is not limited by the issues regarding secondary electrons. The electron beam optics can be optimized for higher current densities (Eq. #eq SEM current density) at higher energies compared to SEM.

To achieve optimal imaging conditions for the thin TEM samples, the working distance has been made short. In most TEMs, the space for the sample holder is only about (5 mm)³ between the two objective lenses for the incoming and transmitted beam. Before reaching a CCD camera, the transmitted beam is sent through several magnification lenses to achieve the high magnification (500.000X is not unusual).

The image formation in TEM can be based on several principles, but practically all images used in this work were made by phase contrast imaging, here called High Resolution TEM or HRTEM. At sufficiently high brightness, electron sources can produce coherent electron beams due to the point-like emitter surface area and small energy spread [8]. The coherent electron beam can be considered as a spherical wave propagating from the emitter and out through the electron optical system, much like a laser beam would propagate through an optical system.

The HRTEM images are often based on the interference of the electron wavefront after it has passed through the sample and reaches a CCD detector to give a phase contrast image of the sample. The image will have a resolution determined of course by the wavelength of the electrons (Eq. #eq SEM de broglie wavelength) but mainly by the imperfections of the electron optics which also perturbs the wavefront. The optimal imaging condition is for a sample thickness about  , where the wavefront is only slightly perturbed by passing through the sample. TEM instruments are normally easily capable of resolving individual shells of a carbon nanotubes. The fine-tuning of the electron optical system to the required resolution can be achieved in about 30 min for many microscopes.

 
TEM images of the same nanostructure using standard 'bright field' TEM vs HAADF STEM. The sample is a gold nanoparticle containing environmental electron beam deposited rod.

Electron Holography

edit

In special TEM microscopes, the diffracted beam can be combined with a part of the original electron beam from the electron gun, and the image that is recorded is an interference pattern that depends on how much the phase of the diffracted beam was changed. By recording such images, one can measure how the electron wave function changes as it passes through or nearby a nanostructure - and this allows you to measure the electric and magnetic fields surrounding nanostructures.

Electron Tomography

edit

By recording numerous TEM images of an object at many different angles, these images can in a computer be combined to create a three-dimensional model of the object. The technique is time consuming but allows you to see nanostructures in 3D.

References

edit
  1. J. Goldstein, D. Newbury, P. Echlin, D. C. Joy, A. D. Romig, C. E. Lyman, C. Fiori, and E. Lifshin. Scanning Electron Microscopy and X-Ray Microanalysis, 2nd Ed. Plenum Press, 1992.
  2. P. J. Goodhew and F. J. Humphreys. Electron Microscopy and Analysis, 2rd Ed. Taylor and Francis, 1988.
  3. S. Humphries. Charged Particle Beams. John Wiley and Sons, 1990. PDF version available at http://www.fieldp.com/cpb/cpb.html.
  4. P. W. Hawkes and E. Kasper. Principles Of Electron Optics. Academic Press, 1989.
  5. L. Reimer. Transmission electron microscopy: Physics of image formation and microanalysis, 3rd Ed. Springer-Verlag, 1993.
  6. P. J. Goodhew and F. J. Humphreys. Electron Microscopy and Analysis, 2rd Ed. Taylor and Francis, 1988.
  7. J. Goldstein, D. Newbury, P. Echlin, D. C. Joy, A. D. Romig, C. E. Lyman, C. Fiori, and E. Lifshin. Scanning Electron Microscopy and X-Ray Microanalysis, 2nd Ed. Plenum Press, 1992.
  8. P. W. Milonni and J. H. Eberly. Lasers. John Wiley & Sons, Inc., 1988.